Jenway 73 Series PC software Bedienungsanleitung

Stöbern Sie online oder laden Sie Bedienungsanleitung nach Ausrüstung Jenway 73 Series PC software herunter. Jenway 73 Series PC software User Manual Benutzerhandbuch

  • Herunterladen
  • Zu meinen Handbüchern hinzufügen
  • Drucken
  • Seite
    / 67
  • Inhaltsverzeichnis
  • LESEZEICHEN
  • Bewertet. / 5. Basierend auf Kundenbewertungen
Seitenansicht 0
Operating Manual
REV A/06-10
73 Series
Spectrophotometer PC Software
Seitenansicht 0
1 2 3 4 5 6 ... 66 67

Inhaltsverzeichnis

Seite 1 - 73 Series

Operating Manual REV A/06-10 73 Series Spectrophotometer PC Software

Seite 2

10 The model number, accessory fitted and wavelength range of the spectrophotometer will be shown in the device information box. If the

Seite 3

11 2.3 Register Register your software/instrument by clicking on Help on the toolbar at the top of the screen. Click on Register and if the comput

Seite 4

12 SECTION 3 – PHOTOMETRICS To open the photometrics measurement mode click on the photometrics icon on the toolbar down the left-hand side of t

Seite 5 - SECTION 1 - INSTALLATION

13 3.1 Settings To set the measurement parameters click on the Settings button. This screen allows the batch name, wavelength and resolut

Seite 6

14 3.1.2 Setting the wavelength To select the wavelength that the zero calibration and sample measurements will be performed at type

Seite 7

15 3.1.4 Auto Logging Auto logging allows repeat measurements of the same sample to be performed with a set time interval between ea

Seite 8 - SECTION 2 – GETTING STARTED

16 To continue click Yes, to cancel click No or Cancel. If the wavelength is changed then the zero calibration must be performed agai

Seite 9

17 If the same sample needs to be measured more than once click on the Auto Log button. The instrument will perform the required nu

Seite 10

18 SECTION 4 – Concentration To open the concentration measurement mode click on the concentration icon on the toolbar down the left-hand side of

Seite 11

19 4.1 Settings To set the measurement parameters click on the Settings button. This screen allows the batch name, wavelength, resolution,

Seite 13

20 4.1.1 Batch name The batch name entered here will be reflected on the name tab at the bottom of the screen and under sample name

Seite 14

21 4.1.3 Setting the resolution The resolution of the absorbance and % transmittance reported is selectable between 0, 0.0, 0.00 or 0.000

Seite 15

22 4.1.5 Entering a factor To select the factor to be used type the factor value into the text box. Factor values of 0.001 to 10,000

Seite 16

23 4.1.7 Auto Logging Auto logging allows repeat measurements of the same sample to be performed with a set time interval between ea

Seite 17

24 wavelength. However a warning will be displayed to warn that critical settings are being changed after the zero calibration has been performed.

Seite 18

25 If the same sample needs to be measured more than once click on the Auto Log button. The instrument will perform the required n

Seite 19

26 Click on the Standard button. The instrument will take a reading and calibrate to the standard concentration. Once the calibration i

Seite 20

27 Insert the cuvette containing the sample into the sample chamber and close the instrument lid. Click on the Read button; the instrum

Seite 21

28 SECTION 5 – SPECTRUM To open the spectrum measurement mode click on the spectrum icon on the toolbar down the left-hand side of the screen.

Seite 22

29 5.1 Settings To set the measurement parameters click on the Settings button. This screen allows the sample name, start and end wavelength

Seite 23

3 Contents Contents...3 SECTION 1

Seite 24

30 5.1.1 Sample name The sample name entered here will be reflected on the name tab at the bottom of the screen.

Seite 25

31 5.1.2 Setting start and end wavelengths This function enables the start and end wavelengths of the spectrum scan to be set. To s

Seite 26

32 5.1.4 Setting the resolution The resolution of the absorbance or % transmittance reported is selectable between 0, 0.0, 0.00 or 0.

Seite 27

33 5.1.6 Auto Logging Auto logging allows repeat measurements of the same sample to be performed with a set time interval between

Seite 28 - SECTION 5 – SPECTRUM

34 At continue click Yes, to cancel click No or Cancel. If the wavelength is changed then the baseline must be performed again at the new

Seite 29

35 The spectrum measurement mode displays the result both graphically and in tabular format. It is possible to zoom in on areas of interest on the

Seite 30

36 5.4.2 Peaks and Valleys Table The peaks and valleys table can be viewed by clicking on the Analysis button on the toolbar at the top of the pa

Seite 31

37 5.4.3 Spectral points analysis The analysis area on the right-hand side of the screen allows points of interest to be selected

Seite 32

38 SECTION 6 - QUANTITATION To open the quantitation measurement mode click on the quantitation icon on the toolbar down the left-hand

Seite 33

39 tab displays all the results, the curve tab displays the current standard curve.

Seite 34

4 6.1.3 Selecting concentration units... 42 6.1.4 Settin

Seite 35

40 6.1 Settings To set the measurement parameters click on the Settings button. This screen allows the sample name, wavelength, concentra

Seite 36

41 6.1.1 Sample name The sample name entered here will be reflected on the name tab at the bottom of the screen and under sample name

Seite 37

42 6.1.2 Setting the wavelength To select the wavelength that the zero calibration and sample measurements will be performed at type th

Seite 38 - SECTION 6 - QUANTITATION

43 6.1.4 Setting the resolution The resolution of the absorbance or % transmittance and concentration reported is selectable between 0,

Seite 39

44 6.1.6 Selecting number of standards The number of standards used to create a standard curve can be changed from 2 to 6 standards. To

Seite 40

45 6.1.8 Auto Logging Auto logging allows repeat measurements of the same sample to be performed with a set time interval between ea

Seite 41

46 6.2.1 Measuring Standards Insert the cuvette containing the blank sample into the sample chamber and close the instrument lid. Cl

Seite 42

47 Once the standards have been measured click on Show Curve to view the quantitation curve. The curve fit can be adjusted in this screen. Click o

Seite 43

48 To continue click Yes, to cancel click No or Cancel. If the wavelength is changed then the zero calibration must be performed again before a s

Seite 44

49 If the settings menu is re-entered following sample measurement it is not possible to change the wavelength. A new quantitation t

Seite 45

5 SECTION 1 - INSTALLATION Insert the Jenway 73 Series software installation CD. The CD should auto run, if not open My Computer,

Seite 46

50 SECTION 7 – Time Scan Kinetics To open the time scan kinetics measurement mode click on the time scan kinetics icon on the toolbar

Seite 47

51 7.1 Settings To set the measurement parameters click on the Settings button. This screen allows the sample name, wavelength, kinetics run time

Seite 48

52 7.1.1 Sample name The sample name entered here will be reflected on the name tab at the bottom of the screen and under sample name

Seite 49

53 7.1.2 Setting the wavelength To select the wavelength that the zero calibration and sample measurements will be performed at type

Seite 50

54 7.1.4 Selecting time interval The time interval that the kinetics measurements are recorded at is selectable between 1, 5, 10,

Seite 51

55 7.1.6 Selecting concentration units The units of concentration can be selected from a number of options: no units, %, ppm, EBC,

Seite 52

56 7.1.8 Setting lag time or start on level In this measurement mode it is possible to delay the commencement of the kinetics measurem

Seite 53

57 7.1.9 Auto Logging Auto logging allows repeat measurements of the same sample to be performed with a set time interval between each

Seite 54

58 To continue click Yes, to cancel click No or Cancel. If the wavelength is changed then the zero calibration must be performed again before a s

Seite 55

59 It is possible to zoom in on areas of interest on the graph by clicking on the graph and holding down the left mouse key then dragging the mouse

Seite 57

60 SECTION 8 - SAVING AND OPENING METHODS In all of the measurement modes it is possible to save the method parameters. There is no limit on the

Seite 58

61 SECTION 9 – SAVING AND OPENING RESULTS There is no limit on the number of results which can be saved. 9.1 Saving results In any of the measur

Seite 59

62 The saved result will be displayed. Only the settings and post measurement tools will be active. The Zero, Read and Auto Log buttons

Seite 60

63 To select the required option put a tick in the box adjacent to the option. To remove an option un-tick the box adjacent to the option.

Seite 61

64 SECTION 12 – ACCESSORIES When the PC software is connected to the instrument it will communicate with the instrument to find out if there is

Seite 62 - SECTION 11 – PRINTING

65 To open the control for the accessory click on the Accessories tab on the toolbar at the top of the screen and click Sipper accessory. To

Seite 63

66 pump direction is reversed by putting a tick in the box adjacent to reverse direction this will send the sample back to the sample container. W

Seite 64 - SECTION 12 – ACCESSORIES

67 Once the required temperature is reached click on the Close button to close the control. The peltier will remain at the set temperature until an

Seite 65

7 Once the software is installed click Close and the software installation is complete.

Seite 66

8 SECTION 2 – GETTING STARTED 2.1 Connection Ensure that the PC and the spectrophotometer are connected via the interface cable, open

Seite 67

9 When an instrument is connected this screen will be shown. Click on Connect to open communication between the PC and the spectroph

Kommentare zu diesen Handbüchern

Keine Kommentare